Laboratory Clinic Open System Automated Blood Chemistry Analyzer

Model: MCL-CA02
Description:
MOQ: 1 set
Lead time: 3-15 working days

Laboratory Clinic Open System Automated Blood Chemistry Analyzer

Overall Performance

Equipment type        Fully automatic discrete, STAT priority

Analysis rate            Constant speed 300T/H(without ISE)

Test principle           Colorimetry, turbidimetry

Analysis method       End-point, 2 kinetics, fixed-time, etc. support single/double wavelength,

linear and non-linear calibration

Sample System

Sample Tray              60 sample positions, random access of all samples

Sample Volume          2-50ul, 0.1ul increment

Sample probe             Liquid level detection, the remaining detection, collision protection etc.

Sample Containers      Sample cups, Micro tube, Original blood-collection tube, etc.

Reagent System 

Reagent Tray               45 reagent positions

Reagent Volume           R1:25-300ul, R2:10-150ul, 0.1ul increment

Reagent Probe              Liquid level detection, the remaining detection, collision protection etc.

Reagent Refrigeration    4-15°C, independent power supply, 24-hour non-stop refrigeration

Reaction System

Reaction Tray          Rotating tray, 120 cuvettes

Cuvette                   Reusable high quality cuvette, Optical diameter 6mm

Reaction Volume      180-500ul

Incubation system     Bi-directional liquid-air device

Temperature             37°C±0.1°C with real-time display

Optical System

Light source                 12V35W halogen lamps, over 2000 hours

Photoelectron road       after spectrophotometry

Wavelength                  340,405,450,510,546,578,630,700nm

Absorbance range         0-3.500Abs

Resolution                    0.0001Abs

Operating System

PC operating system         Windows XP or above

Analysis control software   English version graphical operating software

Report printing                 Report formats support the user-defined mode, QC and state information etc.

System connection            RS-232

Component Details

1. Cuvette

2.  Needle

3. Reagent Tray

4. Sample & reagent Tray

5. Self washing system of the biochemistry analyzer

Working Conditions

Power supply:                 AC 220V±22V, 50Hz±1Hz, power 1.5KVA

Ambient Temperature:      15-30 °C

Humidity :                        35-80%

Water consumption:          7L/Hour

Dimension:                       Bench top: 750mm*520mm*600mm

Weight:                            90KG